GEG XRCT System
& Scanning Services

Customized X-ray Computed Tomography system at ETH Zurich designed for in-situ reactive flow-through experiments, high-resolution micro-CT imaging, and advanced 3D digital rock analysis.

XRCT Scanner in Room NW B86
Custom XRCT Scanner with Rotating Gantry (Room NW B86)

Welcome to the GEG XRCT System and Services

The GEG XRCT system of ETH Zurich, designed for in-situ reactive flow-through experiments, provides resources and services in X-ray CT imaging as well as direct access to a state-of-the-art CT scanner, dedicated software, and in-situ imaging equipment.

GEG Research

Serves cutting-edge experimental research of the GEG Group in geothermal energy, geological CO₂ storage, and hydrogeology.

Academic Collaborations

Supports collaborative scientific projects between GEG and internal ETH departments as well as national and international research groups.

Commercial Services

Provides professional contract scanning services for university research groups, research institutes, and industrial partners.

XRCT Instrument Specifications

Custom-built within a temperature-controlled ETHZ shielding enclosure in Room NW B86.

The customized XRCT scanner is equipped with an adjustable X-ray tube operating up to 130 kVp with a focal spot size of 5 μm at a power of 4–39 W, coupled with a CMOS flat panel detector featuring a structured CsI scintillator of 2940 × 2304 pixels.

Depending on the Field of View (FOV), the scanner delivers a nominal scanning resolution down to 2.5 μm. The variable FOV spans from 7 to 90 mm, outputting 2D image slices from 512×512 up to 8192×8192 pixels.

Air-Bearing Rotating Gantry Design

One of the novel design breakthroughs is a horizontally-rotating gantry where the X-ray tube and detector rotate around the stationary sample. Supported by an ultra-smooth air bearing, this enables 180° and 360° rotation for centered or offset multi-stack scanning.

Samples mounted along the vertical Z-axis can extend up to a scanning length of 400 mm for calibrated, seamless helical scanning under high-P / high-T flow conditions.

LocationETH Zurich Campus Zentrum, Room NW B86
X-Ray Tube VoltageUp to 130 kVp (power 4 – 39 W)
Focal Spot Size5 μm
Detector TypeCMOS Flat Panel with structured CsI scintillator
Detector Matrix2940 × 2304 pixels
Nominal ResolutionDown to 2.5 μm
Field of View (FOV)7 to 90 mm (variable)
Rotation GantryAir-bearing supported, 180° / 360° horizontal rotation
Max Vertical Scan Range400 mm (Z-axis helical scanning)
ReconstructionGPU-accelerated volumetric reconstruction

Software Stack & Digital Rock Physics

SCANCO MicroCT Analysis Software

Dedicated measurement, acquisition, 3D evaluation, visualization, and raw data management suite from SCANCO Medical AG. Provides full control over tomographic calibration, ring artifact reduction, beam hardening corrections, and volumetric slicing.

Thermo Fisher PerGeos & Petrophysics

Industrial-grade digital rock analysis package featuring 3D CT image assembly, segmentation, grain/pore shape characterization, fracture dip/strike analysis, and direct numerical calculations of rock permeability, electrical resistivity, thermal conductivity, and molecular diffusivity.

Funding & Acknowledgements

We gratefully acknowledge financial and technical support from our founding partners and sponsors:

  • SNF REquip Grant Nr. 177031 (Swiss National Science Foundation)
  • Werner Siemens Foundation (Werner Siemens-Stiftung)
  • ETH Zurich Startup Grant
  • SCANCO Medical AG (Technical Support)